Comparison between Microwave Radar Gauge and Pressure Gauge in Monitoring Sea Level Surface at Alexandria Naval Port

Document Type : Original Article

Authors

1 Survey Research Institute (SRI), National Water Research Centre

2 Giza High Institute of Engineering &Technology

Abstract

In this study, a comparison between the observations of the PG and RG were conducted by collecting data from both devices over a period of fourteen months from January 2020 to February 2021.
A sea level monitoring station was established in the Alexandria Naval Port by the National Water Research Centre (NWRC) in collaboration with the Egyptian Military Survey Department of the Ministry of Defense. The microwave radar device was installed next to the previously installed pressure gauge, and data were collected from both devices. By analyzing 20,400 observations collected every 30 minutes over a period of 14 months and conducting T-test and F-test statistical analysis after applying 3ϭ rule filter, as well as representing graphical plots of the observations of both devices at different time periods, the results indicated a somewhat close accuracy between the two devices over long periods of the year. However, the pressure gauge remained stable, and its observations were more consistent than those of the microwave radar device.

Keywords

Main Subjects


Volume 7, Issue 2
Special Issue, Selected papers from the Third International Conference on Advanced Engineering Technologies for Sustainable Development ICAETSD, held on 21-22 November 2023
2024
Pages 1-9